UET System · UET-LAF

LAF HEPA filter and air velocity monitoring device

LAF HEPA & AIR CONTROL monitors differential pressure in laminar-flow and HEPA-filter applications. It calculates filter loading from configured clean and dirty filter pressures; this is not a direct particle or microbiological cleanliness measurement.

Measurement and integration scope

  • Active measurements are differential pressure and optional air velocity.
  • Filter loading is calculated from configured pressure references.
  • Three filter alarm levels and optional air velocity alarms can be configured.
  • The optional SAV.P probe is read over RS485 Modbus RTU with the device acting as master; analog outputs are unused in this firmware variant.
LAF HEPA filter and air velocity monitoring device

Selection and commissioning

Commission filter reference pressures, alarm thresholds, probe selection and placement for the application. A filter-loading indication does not replace HEPA integrity testing, airflow validation or cleanroom classification.

Technical documentation

Technical descriptions are based on the existing model manual. Check the ordered hardware and firmware revision against the current manual.

User manual (PDF, Turkish)

Related products and applications

Discuss your technical requirements